Automated CT Lung Cancer Screening Workflow using 3D Camera

Auteurs : Brian Teixeira, Vivek Singh, Birgi Tamersoy, Andreas Prokein, Ankur Kapoor

Accepted at MICCAI 2023

Résumé : Despite recent developments in CT planning that enabled automation in patient positioning, time-consuming scout scans are still needed to compute dose profile and ensure the patient is properly positioned. In this paper, we present a novel method which eliminates the need for scout scans in CT lung cancer screening by estimating patient scan range, isocenter, and Water Equivalent Diameter (WED) from 3D camera images. We achieve this task by training an implicit generative model on over 60,000 CT scans and introduce a novel approach for updating the prediction using real-time scan data. We demonstrate the effectiveness of our method on a testing set of 110 pairs of depth data and CT scan, resulting in an average error of 5mm in estimating the isocenter, 13mm in determining the scan range, 10mm and 16mm in estimating the AP and lateral WED respectively. The relative WED error of our method is 4%, which is well within the International Electrotechnical Commission (IEC) acceptance criteria of 10%.

Soumis à arXiv le 27 Sep. 2023

Explorez l'arbre d'article

Cliquez sur les nœuds de l'arborescence pour être redirigé vers un article donné et accéder à leurs résumés et assistant virtuel

Accédez également à nos Résumés, ou posez des questions sur cet article à notre Assistant IA.

Beta

Recherchez des articles similaires (en version bêta)

En cliquant sur le bouton ci-dessus, notre algorithme compare le titre et le résumé de cet article à tous les articles de la même catégorie arXiv que nous avons résumés, et trouve les plus proches. Vous pouvez le relancer de temps en temps : les résultats s'améliorent à mesure que notre base grandit.